Încercaţi să căutaţi în funcţie de cuvinte-cheie, codul produsului sau codul de comandă ori numărul de serie, de exemplu, „CM442” sau „Informaţii tehnice”
Introduceţi cel puţin 2 caractere pentru a începe căutarea.
Liquiline System CA80NO offers precise, regulation-compliant online measurement of nitrite. It supports you in meeting the specified limits in drinking water, mineral water or food production. Like all Liquiline System analyzers, it enables plug & play of up to four Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating and maintenance costs. Advanced diagnostics with remote access ensure process safety.
Liquiline System CA80NO supports you in keeping the strict nitrite limits for:
Drinking water
Mineral water
Raw water for food production
Avantaje
Compliance with standard colorimetric measuring principle - naphthylamine method - following ISO 6777 and DIN EN 26777 ensures comparability to lab measurements.
Reduced operating costs through automatic calibration and cleaning.
Simple maintenance: no tools required.
Advanced diagnostics with remote access for increased process safety.
Fast commissioning and plug & play thanks to Memosens technology and user-friendly operation.
Easy upgrade of functionality - even to a complete measuring station - simply by adding modules and connecting Memosens sensors. Reduces installation effort.
Drive connectivity:Liquiline Edge Module CYY7 transfers process data to Netilion, independently of the fieldbus. This allows remote overview and the use of IIoT services. It’s available in new products or as upgrade to installed Liquiline platforms.
Make life easier, increase process safety and stay connected with Memosens 2.0 digital technology.
Liquiline: cutting-edge transmitter platform
By choosing the Liquiline platform, you’re effectively choosing to spend less money on stock holding, save installation time and gain operational safety!
The monitoring solution provides the Visp fish farm with stable, reliable oxygen control, enabling optimal conditions for fish and reducing effort and risk.
Proline Teqwave MW 300 delivers stable real-time data taken directly from the process, reduces maintenance work, and allows quicker intervention if changes to the process are required.
Kerry DCF produces spreads and butter products for the retail market. The company wanted to automate its test for chemical oxygen demand in its treated effluent to save time and optimise the process.
Bonumose decided to enhance, automate and digitalize its tagatose sweetener manufacturing process with Endress+Hauser instrumentation to reach its company growth goals
Housing (open version): 793 x 530 x 417 mm 31.22 x 20.87 x 16.42 in Housing (closed version): 793 x 530 x 463 mm 31.22 x 20.87 x 18.23 in Housing with base: 1723 x 530 x 463 mm 67.83 x 20.87 x 18.23 in
Design
Open design, cabinet and stand housing High-Performance plastic ASA-PC, additional stand coated steel
Process temperature
4 to 40 °C (39 to 104 °F)
Ambient temperature
5 to 40 °C (41 to 104 °F) Outdoor version: - 20 to 40 °C (-4 to 104 °F)
Easy upgrade to measuring station with up to four digital Memosens sensors Automatic calibration and cleaning User-configurable measuring, cleaning and calibration intervals Optional cooling module for standard solution 2 channel version optional User-definable measuring ranges Modular design for easily extensible functionality
Application
Limit value monitoring of nitrite for potable water and mineral water applications
Power supply
100 to 120 VAC / 200 to 240 VAC ± 10% 24 VDC ± 10% 50 ± 1 or 60 ± 1,2 Hz
Output / communication
2x 0/4 to 20 mA Modbus RS485/TCP (optional) Ethernet (optional) Alarmrelay
Input
1 or 2 measuring channel 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measuring range
10 to 500 µg/l NO2-N 0.2 to 3.0 mg/l NO2-N 0.1 to 1.0 mg/l NO2-N 0.1 to 1.0 mg/l with dilution function to maximum 5 to 50 mg/l NO2-N
Consumables
Reagents and standard solutions CY80NO as well as cleaner CY800 are necessary for the operation Regular maintenance is done with the parts of the maintenance kit CAV800
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex
Ensure optimal CA80 analyzer performance with expert commissioning and first year preventive maintenance service for seamless operation and reduced downtime